Conventional positive pressure ventilation
IPPV via ETT with continuous flow
of heated and humidified gas allows the non-paralysed baby to breathe
spontaneously. The ventilator is time-cycled, pressure limited (TCPL) where the
user sets the positive inspira-tory pressure (PIP), inspiration time (Ti) and
ventilator rate. In this mode the tidal volume is determined by the lung
compliance. Some ventilators can adjust PIP within a set range to deliver a set
tidal volume (volume guarantee). Some ventilators can terminate inspiration
when a set volume is reached or when inspiratory flow is declining below a
threshold level. Whichever method is chosen the user must be familiar with the
operation and limitations of the ventilator.
•
Worsening
respiratory failure, e.g. RDS.
•
Impending
or actual respiratory arrest from any cause.
•
Recurrent
apnoeas.
•
Massive
pulmonary haemorrhage.
•
Severe
cardiac failure.
•
Persistent
pulmonary hypertension of the newborn.
• Severe congenital lung
malformation, e.g. diaphragmatic hernia.
• Severe HIE.
• Anaesthesia.
•
Peak
inspiratory pressure (PIP).
•
PEEP.
•
TI and
expiratory (TE) time (often expressed as I:E ratio).
•
Inspired
O2% or fraction inspired O2 (FiO2).
•
Gas
flow (L/min) through ventilator circuit (may not be adjustable).
•
Review
and adjust ventilation settings soon after commencement.
•
Monitor
blood gases and adjust ventilation as appropriate. Acceptable limits will
depend on the clinical situation, however, as a guide in preterm infants; pH
7.2–7.35, PCO2 5–8kPa, PO2 6–10kPa, saturation 90–95%,
expired tidal volume around 5mL/kg.
•
If PaO2 is too low—i FiO2, or ‘rise’mean airway pressure (the latter
by either ‘rise’PIP, ‘rise’PEEP, or
‘fall’ TE which will ‘rise’rate as TI stays constant). Do the opposite if PaO2
is too high.
If
PaCO2 is too high—i alveolar
ventilation, i.e. minute volume, by
‘rise’PIP, or ‘fall’ PEEP, or ‘rise’rate. Do the opposite if PaCO2
is too low.
May present as systemic collapse, ‘fall
fall’ PaO2, or ii PaCO2. Ventilate with manual system,
e.g. T-Piece (preferably with PEEP), and O2 as required. Rapid
improvement suggests ventilator problem. Otherwise consider ob-structed ETT,
displaced ETT, pneumothorax, or non-respiratory disease, e.g. intraventricular
haemorrhage (IVH), gut perforation.
May present as slow deterioration
in overall clinical condition, ‘rise’PaCO2, or ‘fall’ PaO2.
Consider: worsening respiratory
disease; partial ETT obstruc-tion; airway circuit leak; non-respiratory
disease.
As condition starts to improve aim
to wean ventilation. Wean O2 to low-est needed to maintain adequate
PaO2 (d retinopathy risk). As lung com-pliance improves wean PIP to
maintain appropriate expired tidal volume (d risk of pulmonary air leak) in
2cmH2O steps until 12–14cmH2O (moni-tor blood gases).
Then wean rate by 5 or 10 increments until 10–20breaths/ min. Following
extubation it is often helpful in preterm infants to start nasal CPAP 5cmH2O.
Extubation without CPAP may be appropriate after short-term ventilation.
Related Topics
Privacy Policy, Terms and Conditions, DMCA Policy and Compliant
Copyright © 2018-2023 BrainKart.com; All Rights Reserved. Developed by Therithal info, Chennai.