Home | | Pharmaceutical Drug Analysis | Preparation of Thin Layers on Plates - Thin Layer Chromatography (TLC)

Chapter: Pharmaceutical Drug Analysis: Thin Layer Chromatography (TLC)

Preparation of Thin Layers on Plates - Thin Layer Chromatography (TLC)

Preparation of Thin Layers on Plates - Thin Layer Chromatography (TLC)
The paramount importance with regard to the preparation of thin layer is that it must be uniform and consistent throughout.

PREPARATION OF THIN LAYERS ON PLATES

 

The paramount importance with regard to the preparation of thin layer is that it must be uniform and consistent throughout. Various means have been put forward to apply thin layers of powdered or their suspen-sions or their slurries to the carrier plates with a view to achieve an uniform layer throughout the length of the plates. These are namely :

 

(a) Pouring of Layers : 

In order to obtain layers of equal thickness, a measured amount of the suspen-sion or slurry is placed on a given-size plate that is rested on an absolutely labelled surface. The plate is subsequently tipped backward and forward to permit the slurry (or suspension) to spread uniformly on the surface of the plate.

 

(b) Dipping : 

Peifer* in 1962, was pioneer in introducing this technique, whereby two plates at a time back-to-back are dipped together in a slurry of the adsorbent in either chloroform or chloroform-methanol. However, this particular methods is not much in use now-a-days.

 

(c) Spraying : 

Reitsema** was first to develop this method by making use of a small paint-sprayer for the distribution of the suspension or slurry onto the surface of the glass-plate.

 

Disadvantages : There are mainly two major disadvantages of this technique, namely :

 

(i) Non-uniformity of layers on a single-plate, and

 

(ii) Variation observed from one plate to the other was significant.

 

Belgian Patent No : 625012 : It essentially consists of spraying either molten or partially molten absorbent onto a glass plate, for instance : an alumina film prepared by melting and aluminium rod with an oxyacetylene flame and subsequently spraying the molten adsorbent onto a glass plate.

(d) Spreading : 

In this particular case, the suspension or slurry is put in an ‘applicator’, which is subsequently moved either over the stationary glass-plate or vice-versa i.e., it is held stationary while the glass plate is pulled or pushed through. This technique termed as ‘spreading’ usually yields uniform thin layers on the glass plates.

 

Kirchner’s* technique essentially consists of :

 

·              selecting uniform surfaced glass plates,

 

·              placing them between glass or metal guides which are thicker than glass plates by the amount that is desired for the adsorbent layer, and

 

·              spreading the slurry on the glass plate with the help of a glass rod.

 

Egon Stahl’s apparatus exclusively designed for the application of thin-adsorbent layers which broadly comprises of two major parts, namely :

 

(i) Aligning Tray : It is a tray on which the glass plates are placed in a series or in-a-line, and

 

(ii) Spreader : It holds the spreading mixture (as a slurry or suspension) and applies it uniformly in a thin-layer.

 

Figure 28.1 (a) shows the operation of thin-layer spreader ; while Figure 28.1 (b) depicts the aligning tray with glass-slide partially coated.


In Figure 28.1 (a)                  

 

A = Glass plate,

 

B = Spreader, and

 

C = Slurry of the adsorbent.

 

Here, the slurry (C) is put in the spreader (B) and then moved along the direction of application onto the surface of the glass plate (A) to obtain an uniform layer,

 

In Figure 28.1 (b)                  

 

A = Glass plate,

 

B = Layer of adsorbent, and

 

C = Aligning tray

 

The slurry of the adsorbent is introduced into the aligning tray (C) which is then moved onto the glass plate (A) to obtain an uniform layer of adsorbent (B).

 

In 1962, a well known West German firm DESAGA introduced a much improved and simplified ver-sion of a TLC-spreader that could conveniently give uniform layer thickness raging from 0.00 to 2,0 mm.

 

(e) ‘TLC-Plates ready-for Use’ (or Pre-coated Plates)

 

E. Merck AG (Darmstadt, West Germany) was pioneer in introducing two types of ready-for-use TLC plates either on glass or polysheets (ethyleneterephthalate), namely :

 

(i) Ready-for-use TLC Plates with Cellulose-F, and

 

(ii) Ready-for-use TLC Plates with Silica Gel F-254.

 

Interestingly, these precoated TLC-plates essentially have : first, a special abrasive-resistant layer con-taining no gypsum ; and secondly, the layer contains a reliable fluorescent indicator that is excited to emit a fluorescence under either a short-wave or a long-wave UV light.

 

Advantages : These ‘ready-for-use’ TLC-plates have the following advantages namely :

 

·              It may be safely activated at 110-120° C, before, use,

 

·              The properties of the layer minimise spot-diffusion that helps both more strong concentration of spots and more distinctive separations with higher sensitivity,

 

·              It may accept more corrosive spray-reagents, for example : conc. sulphuric acid, phosphoric acid, phosphomolybdic acid, perchloric acid on antimony trichloride ; and the sprayed plates could be heated upto 110-120 °C without any darkening whatsoever,

 

·              The migration rate is slightly enhanced when compared to hand coated plates, and

 

·              The TLC plates may be cut into strips by the aid of a glass cutter applied on the reverse side.

 