Home | | Anesthesiology | Anesthesia Ventilators

Chapter: Clinical Anesthesiology: Anesthetic Equipment & Monitors : The Anesthesia Machine

Anesthesia Ventilators

Anesthesia Ventilators
Ventilators are used extensively in the operating room (OR) and the intensive care unit (ICU).

VENTILATORS

Ventilators are used extensively in the operating room (OR) and the intensive care unit (ICU). All modern anesthesia machines are equipped with a ventilator. Historically OR ventilators were simpler and more compact than their ICU counterparts. This distinction has become blurred due to advances in technology together with an increasing need for “ICU-type” ventilators as more critically ill patients come to the OR. The ventilators on some modern machines are just as sophisticated as those in the ICU and have almost the same capabilities. After a general discussion of basic ventilator principles, this section reviews the use of ventilators in conjunction with anesthesia machines.

Overview

Ventilators generate gas flow by creating a pres-sure gradient between the proximal airway and the alveoli. Older units relied on the generation of nega-tive pressure around (and inside) the chest (eg, iron lungs), whereas modern ventilators generate posi-tive pressure and gas flow in the upper airway.

Ventilator function is best described in rela-tion to the four phases of the ventilatory cycle: inspiration, the transition from inspiration to expiration, expiration, and the transition from expiration to inspiration. Although several clas-sification schemes exist, the most common is based on inspiratory phase characteristics and the method of cycling from inspiration to expiration.

Other classification categories may include power source (eg, pneumatic-high pressure, pneumatic-Venturi, or electric), design (single-circuit system, double-circuit system, rotary piston, linear piston), and control mechanisms (eg, electronic timer or microprocessor).

A. Inspiratory Phase

During inspiration, ventilators generate tidal volumes by producing gas flow along a pressure gradient. The machine generates either a con-stant pressure (constant-pressure generators) or constant gas flow rate (constant-flow generators) during inspiration, regardless of changes in lung mechanics ( Figure 4–23). Nonconstant generators produce pressures or gas flow rates that vary dur-ing the cycle but remain consistent from breath to breath. For instance, a ventilator that generates a flow pattern resembling a half cycle of a sine wave (eg, rotary piston ventilator) would be classified as a nonconstant-flow generator. An increase in airway resistance or a decrease in lung compliance would increase peak inspiratory pressure but would not alter the flow rate generated by this type of ventila-tor (Figure 4–24).




B. Transition Phase from Inspiration to Expiration

Termination of the inspiratory phase can be trig-gered by a preset limit of time (fixed duration), a set inspiratory pressure that must be reached, or a predetermined tidal volume that must be deliv-ered. Time-cycled ventilators allow tidal volume and peak inspiratory pressure to vary depend-ing on lung compliance. Tidal volume is adjusted by setting inspiratory duration and inspiratory flow rate. Pressure-cycled ventilators will not cycle from the inspiratory phase to the expiratory phase until a preset pressure is reached. If a large circuit leak decreases peak pressures significantly, pressure-cycled ventilator may remain in the inspiratory phase indefinitely. On the other hand,small leak may not markedly decrease tidal vol-ume, because cycling will be delayed until the pressure limit is met. Volume-cycled ventilators vary inspiratory duration and pressure to deliver a preset volume. In reality, modern ventilators over-come the many shortcomings of classic ventilator designs by incorporating secondary cycling param-eters or other limiting mechanisms. For example, time-cycled and volume-cycled ventilators usu-ally incorporate a pressure-limiting feature that terminates inspiration when a preset, adjustable safety pressure limit is reached. Similarly a volume-preset control that limits the excursion of the bel-lows allows a time-cycled ventilator to function somewhat like a volume-cycled ventilator, depend-ing on the selected ventilator rate and inspiratory flow rate.

C. Expiratory Phase

The expiratory phase of ventilators normally reduces airway pressure to atmospheric levels or some preset value of positive end-expiratory pressure (PEEP). Exhalation is therefore passive. Flow out of the lungs is determined primarily by airway resistance and lung compliance. Expired gases fill up the bellows; hey are then relieved to the scavenging system. PEEP is usually created with an adjustable spring valve mechanism or pneumatic pressurization of the exhalation (spill) valve.

D. Transition Phase from Expiration to Inspiration

Transition into the next inspiratory phase may be based on a preset time interval or a change in pressure. The behavior of the ventilator during this phase together with the type of cycling from inspi-ration to expiration determines ventilator mode.During controlled ventilation, the most basic mode of all ventilators, the next breath always occurs after a preset time interval. Thus tidal vol-ume and rate are fixed in volume-controlled ven-tilation, whereas peak inspiratory pressure is fixed in pressure-controlled ventilation. Controlled ventilation modes are not designed for spontane-ous breathing. In the volume-control mode, the ventilator adjusts gas flow rate and inspiratory time based on the set ventilatory rate and I:E ratio (Figure 4–25A). In the pressure-control mode, inspiratory time is also based on the set ventilator rate and inspiratory-to-expiratory (I:E) ratio, but gas flow is adjusted to maintain a constant inspira-tory pressure (Figure 4–25B).


In contrast, intermittent mandatory ventila-tion (IMV) allows patients to breathe spontaneously between controlled breaths. Synchronized inter-mittent mandatory ventilation (SIMV) is a further refinement that helps prevent “fighting the ventila-tor” and “breath stacking”; whenever possible, the ventilator tries to time the mandatory mechanical breaths with the drops in airway pressure below the end-expiratory pressure that occur as the patient ini-tiates a spontaneous breath.

Study Material, Lecturing Notes, Assignment, Reference, Wiki description explanation, brief detail
Clinical Anesthesiology: Anesthetic Equipment & Monitors : The Anesthesia Machine : Anesthesia Ventilators |


Privacy Policy, Terms and Conditions, DMCA Policy and Compliant

Copyright © 2018-2023 BrainKart.com; All Rights Reserved. Developed by Therithal info, Chennai.