What is pacemaker hysteresis?
It is generally desirable to preserve intrinsic
cardiac rhythm if the rate and rhythm are adequate to maintain an acceptable
cardiac output. If the heart rate transiently decreases below that to which the
pacemaker is set, it may not be necessary for the pacemaker to discharge. This
is especially important for patients whose intrinsic cardiac rhythm is normal
sinus and who have ventricular pace-makers. Considering the contribution of the
atrial “kick” to cardiac output, the patient may have a better cardiac output
in a slow sinus rhythm than would be created by single-chamber pacing. The
additional delay that the pace-maker allows before initiating discharge after
an intrinsic beat is known as hysteresis.
For example, if the pacemaker is set to pace at
a rate of 60 beats per minute, then the pacing interval is 1,000 msec (1
second) between each paced beat. If we wish to program the pacemaker to wait
until the intrinsic heart rate is less than 50 per minute (1,200 msec between
beats) before firing, then the hysteresis is considered to be 200 msec. Once
1,200 msec have elapsed since the last intrinsic beat, the pacemaker will begin
pacing every 1,000 msec. For the intrinsic rhythm to inhibit the pacemaker, an
intrinsic beat will have to occur within the 1,000 msec interval following the
last paced beat.
Related Topics
Privacy Policy, Terms and Conditions, DMCA Policy and Compliant
Copyright © 2018-2026 BrainKart.com; All Rights Reserved. Developed by Therithal info, Chennai.