Home | | Radiology | Fluoroscopy

Chapter: Basic Radiology : The Physical Basis of Diagnostic Imaging

Fluoroscopy

Fluoroscopy
If, instead of using the light from a fluorescent screen to blacken a film, one viewed the fluorescent screen directly with the naked eye, then one would be performing fluo-roscopy as it was done in the early days of medical x-ray use.

Fluoroscopy

 

If, instead of using the light from a fluorescent screen to blacken a film, one viewed the fluorescent screen directly with the naked eye, then one would be performing fluo-roscopy as it was done in the early days of medical x-ray use. Unfortunately, the image made in this fashion was very dim, even at a high exposure rate to the patient, so modern fluo-roscopy uses an image intensifier that amplifies the light from a fluorescent screen. A typical fluoroscopic imaging system is shown in Figure 2-3. The image intensifier tube is an evacuated glass or metal tube with a fluorescent screen (input phosphor) that glows with the image produced by the x-ray pattern that exits the patient. The light from theinput phosphor causes ejection of electrons from a photo-electric material adjacent to the input phosphor. These elec-trons are accelerated via a high voltage (30 kV), as well as being focused to preserve the image onto a small (1-inch di-ameter) screen (the output phosphor), which glows with the image because of the energy deposited by the impact of the accelerated electrons. The output phosphor glows much more brightly than the input phosphor (about 3000 times) because of the energy gain provided by the acceleration of the electrons and also because of minification of the image on the output phosphor. The image on the output phosphor can be viewed with the naked eye, usually with a series of lenses and mirrors, but the image is more commonly viewed by fo-cusing a video camera onto the output phosphor and viewing the image on a TV monitor via a closed-circuit TV system. The fluoroscopic image generally has less contrast and less resolution of fine detail than a radiographic image; however, it is clearly convenient to view the image in real time— particularly when observing the flow of radiopaque contrast agents ingested or injected into the body. (These contrast ma-terials, such as iodine or barium compounds, have a higher atomic number than soft tissue, hence, absorb more x-rays.) During fluoroscopic examinations, the x-ray tube is typically operated below 100 kV and below 3 mA tube current. Even so, entrance exposure rates (at the point where the x-ray beam enters the patient) are about 2 to 5 R/min, depending on patient thickness; hence, fluoroscopic examinations gen-erally result in significantly higher patient exposures than do radiographic examinations.

 

Fluoroscopic systems generally have an automatic bright-ness control in which the brightness of the output phosphor is sensed by a light detector. The brightness signal from this detector is compared to a reference level, and the difference signal is used to instruct the x-ray generator to vary mil-liamperage or kilovoltage (or both) in order to maintain a constant brightness at the output phosphor. For example, after ingestion of barium in a barium-swallow examination, the barium absorbs significantly more x-rays, and the image would tend to go dark without such a system; however, as the brightness falls below the reference level, the automatic brightness control causes the x-ray generator to increase the milliamperage or kilovoltage to maintain a constant bright-ness on the monitor.

Figure 2-3. Fluoroscopic imaging system.



 

Study Material, Lecturing Notes, Assignment, Reference, Wiki description explanation, brief detail
Basic Radiology : The Physical Basis of Diagnostic Imaging : Fluoroscopy |


Privacy Policy, Terms and Conditions, DMCA Policy and Compliant

Copyright © 2018-2024 BrainKart.com; All Rights Reserved. Developed by Therithal info, Chennai.