Home | | Medical Electronics | DC Defibrillation

Chapter: Medical Electronics : Assist Devices and Bio-Telemetry

DC Defibrillation

To overcome the disadvantage of defibrillation method In 1962,Bernard lawn from Harward school of public health and peter bent of Brigham hospital developed a new method known as dc defibrillation.

DC DEFIBRILLATION

 

INTRODUCTION

 

·            To overcome the disadvantage of defibrillation method In 1962,Bernard lawn from Harward school of public health and peter bent of Brigham hospital developed a new method known as dc defibrillation.

 

·            In this dc defibrillation method , a capacitors charged to a high dc voltage and then rapidly discharged through electrodes across the chest of patient.

 

·            DC defibrillation is capable of correcting both the atrial fibrillation and ventricular fibrillation.

 

·            DC method produces some harm to the patient. Depending on the energy setting in the defibrillator , the amount of electrical energy discharged by the capacitor ranges between 100 to 400 joules. Discharge portion is approximately 5 ms.

 

·            In discharge waveform ,the peak value ofs current is nearly20 A and the wave is monophasic in nature.

 

·            Monophasic means most of the excursion of curve is above the base line.


Energy level of a defibrillator can be controlling:The voltage amplitudeVPof the defibrillator by varying the setting onthe varactor or Duration of the defibrillator pulse. The energy (WA) stored in the capacitor C and available for the defibrillation is:

 

Lown waveform: Curve 1 shows a typical discharge pulse of defibrillator which called ―Lown waveform.

 

I rises rapidly to app. 20 A

 

Then I decays to 0 with 5 ms

 

A negative pulse is produced for 1 to 2 ms

 

The pulse width defined as the time that elapses between the start of the impulse and the moment that the current intensity passes the zero line for the first time and changes direction (5 ms or 2.5 ms)

 


DUAL PEAK DC DEFIBRILLATOR

 

If peak voltage is as high as 6000V is used there is a possibility of damaging myocardium and the chest walls.

 

·           Produce dual peak wavereform of longer duraoltation at lower voltage.

 

·           Effective defibrillation is  achieved in adults with lower level of delivered energy.

·           Energy range is between 50 to 200W-sec or joules.

 

·           Effective defibrillation at the desirable lower voltage levels is also possible with the truncated waveform.

 

·           The amplitude of the waveform is relatively constant, but is varied to get required energy.

 

·           Large electrodes are used for the proper delivery of large current through the surface of the skin.

 

·           These electrodes are called as paddles .

 

EXTERNAL DEFIBRILLATOR:

 

 

·        A unit based on computer technology and designed to analyze the heart rhythm itself, and then advise whether a shock is required.

 

·        It is designed to be used by lay persons, who require little training.

 

·        It is usually limited in their interventions to delivering high joule shocks for VF and VT rhythms

 

·        The automatic units also take time (generally 10-20 seconds) to diagnose the rhythm, where a professional could diagnose and treat the condition far quicker with a manual unit.

 

·        Automated external defibrillators are generally either held by trained personnel who will attend incidents, or are public access units which can be found in places including corporate and government offices, shopping centers, airports, restaurants,

 

·        AEDS require self-adhesive electrodes instead of hand-held paddles for the two following reasons:

 

·        The ECG signal acquired from self-adhesive electrodes usually contains less noise and has higherquality allows faster and more accurate analysis of the ECG better shock decisions. Hands off‖ defibrillation is a safer procedure for the operator, especially if the operator has little or no training.


DC DEFIBRILLATOR WITH SYNCHRONIZER

 

·        Synchronization means, synchronized the working of the heart with the pacemaker. Synchronized DC defibrillator allows the electric shock at the right point on the ECG of the patient.

 

·        Electric shock is delivered approximately 20 to 30 ms after the peak of R wave of patients ECG.

 



Working

 

·  ECG waveform is traced from the patient.

R-wave in the output of ECG amplifier triggers the time delay circuit .It gives the delay of 30 ms approximately. After that, defibrillator circuit is switched ON. So that, the

capacitor    discharges the electric shock to the patient’s heart.

 

·  The moment at which electric shock occurs is noted by producing the marker pulse on monitoring display.

 

·  This type of circuit is preferred in cardiac emergencies

 

·    The sudden cardiac arrest can be treated using a defibrillator and 80 percent of the patients will be cured from the cardiac arrest if the is given within one minute of the attack.

 

Electrodes used for defibrillation

 

·                       These paddles have metal disks of 8 to 10 cm in diameter for external use.

 

·                       For internal use smaller paddles are used on infants and children.

·                       For external use, pair of electrodes are firmly pressed against the patients chest.

 

Need of Insulation Handle

 

·                       To prevent the person applying the electrodes from accidential electric shock specially insulated handles are provided in the paddles.

 

·                       When paddles are properly positioned , this prevents the patient from receiving a shock.

 

·                       In earlier equipment a foot switch is used instead of thumb switch.

 

Need of Thumb Switch

 

·                       There is a possibility of someone accidentally stepping on the foot switch in the excitement of an emergency before the paddles are placed. So thumb switches are mostly preferred.

 

Charging of Defibrillators

 

·                       In some defibrillators charging is done by means of a charge switch located in the front panel of the unit.

 

·                       The charge switch is located in the handle of one of its paddles.

·                       In few defibrillators the charging process begins automatically after discharge.

 

Types of Electrodes

 

Two electrodes are

Ø     Anterior-anterior

 

Ø     Anterior-posterior

 

·                       Anterior-anterior paddles are applied to the chest. Anterior-posterior paddles are applied to both the patients chest wall and back so that energy is delivered through the heart.

 

·                       Specially designed pediatric paddles are available with diameter ranging from 2 to 6 cm.

 

·                       Internal paddles can be either gas-sterilized or autoclaved.

 

Indication Meter

 

·                       Most of the defibrillators include a watt second meter to inducate the amount of energy stored in the capacitor before discharge.

 

·                       The energy indicated on the meter is lost or dissipated as heat in the components inside the unit.

 


Study Material, Lecturing Notes, Assignment, Reference, Wiki description explanation, brief detail
Medical Electronics : Assist Devices and Bio-Telemetry : DC Defibrillation |


Privacy Policy, Terms and Conditions, DMCA Policy and Compliant

Copyright © 2018-2024 BrainKart.com; All Rights Reserved. Developed by Therithal info, Chennai.